blob: ca551ee3e9e98db85660db92b2ea6d6b66766164 [file] [log] [blame]
# MAX30101 heart rate sensor
# Copyright (c) 2017, NXP
# SPDX-License-Identifier: Apache-2.0
menuconfig MAX30101
bool "MAX30101 Pulse Oximeter and Heart Rate Sensor"
depends on I2C && HAS_DTS_I2C
if MAX30101
config MAX30101_SMP_AVE
int "Sample averaging"
range 0 7
default 0
help
To reduce the amount of data throughput, adjacent samples (in each
individual channel) can be averaged and decimated on the chip by
setting this register. Set to 0 for no averaging.
0 = 1 sample (no averaging)
1 = 2 samples
2 = 4 samples
3 = 8 samples
4 = 16 samples
5 = 32 samples
6 = 32 samples
7 = 32 samples
config MAX30101_FIFO_ROLLOVER_EN
bool "FIFO rolls on full"
help
Controls the behavior of the FIFO when the FIFO becomes completely
filled with data. If set, the FIFO address rolls over to zero and the
FIFO continues to fill with new data. If not set, then the FIFO is
not updated until FIFO_DATA is read or the WRITE/READ pointer
positions are changed.
config MAX30101_FIFO_A_FULL
int "FIFO almost full value"
range 0 15
default 0
help
Set the trigger for the FIFO_A_FULL interrupt
choice MAX30101_MODE
prompt "Mode control"
default MAX30101_MULTI_LED_MODE
config MAX30101_HEART_RATE_MODE
bool "Heart rate mode"
help
Set to operate in heart rate only mode. The red LED channel is
active.
config MAX30101_SPO2_MODE
bool "SpO2 mode"
help
Set to operate in SpO2 mode. The red and IR LED channels are active.
config MAX30101_MULTI_LED_MODE
bool "Multi-LED mode"
help
Set to operate in multi-LED mode. The green, red, and/or IR LED
channels are active.
endchoice
config MAX30101_ADC_RGE
int "ADC range control"
range 0 3
default 2
help
Set the ADC's full-scale range.
0 = 7.81 pA/LSB
1 = 15.63 pA/LSB
2 = 31.25 pA/LSB
3 = 62.5 pA/LSB
config MAX30101_SR
int "ADC sample rate control"
range 0 7
default 0
help
Set the effective sampling rate with one sample consisting of one
pulse/conversion per active LED channel. In SpO2 mode, these means
one IR pulse/conversion and one red pulse/conversion per sample
period.
0 = 50 Hz
1 = 100 Hz
2 = 200 Hz
3 = 400 Hz
4 = 800 Hz
5 = 1000 Hz
6 = 1600 Hz
7 = 3200 Hz
config MAX30101_LED1_PA
hex "LED1 (red) pulse amplitude"
range 0 0xff
default 0xff
help
Set the pulse amplitude to control the LED1 (red) current. The actual
measured LED current for each part can vary significantly due to the
trimming methodology.
0x00 = 0.0 mA
0x01 = 0.2 mA
0x02 = 0.4 mA
0x0f = 3.1 mA
0xff = 50.0 mA
config MAX30101_LED2_PA
hex "LED2 (IR) pulse amplitude"
range 0 0xff
default 0x33
help
Set the pulse amplitude to control the LED2 (IR) current. The actual
measured LED current for each part can vary significantly due to the
trimming methodology.
0x00 = 0.0 mA
0x01 = 0.2 mA
0x02 = 0.4 mA
0x0f = 3.1 mA
0xff = 50.0 mA
config MAX30101_LED3_PA
hex "LED2 (green) pulse amplitude"
range 0 0xff
default 0xff
help
Set the pulse amplitude to control the LED3 (green) current. The
actual measured LED current for each part can vary significantly due
to the trimming methodology.
0x00 = 0.0 mA
0x01 = 0.2 mA
0x02 = 0.4 mA
0x0f = 3.1 mA
0xff = 50.0 mA
if MAX30101_MULTI_LED_MODE
config MAX30101_SLOT1
int "Slot 1"
range 0 7
default 3
help
Set which LED and pulse amplitude are active in time slot 1.
0: None (disabled)
1: LED1 (red), LED1_PA
2: LED2 (IR), LED2_PA
3: LED3 (green), LED3_PA
4: None (disabled)
5: LED1 (red), PILOT_PA
6: LED2 (IR), PILOT_PA
7: LED3 (green), PILOT_PA
config MAX30101_SLOT2
int "Slot 2"
range 0 7
default 0
help
Set which LED and pulse amplitude are active in time slot 2.
0: None (disabled)
1: LED1 (red), LED1_PA
2: LED2 (IR), LED2_PA
3: LED3 (green), LED3_PA
4: None (disabled)
5: LED1 (red), PILOT_PA
6: LED2 (IR), PILOT_PA
7: LED3 (green), PILOT_PA
config MAX30101_SLOT3
int "Slot 3"
range 0 7
default 0
help
Set which LED and pulse amplitude are active in time slot 3.
0: None (disabled)
1: LED1 (red), LED1_PA
2: LED2 (IR), LED2_PA
3: LED3 (green), LED3_PA
4: None (disabled)
5: LED1 (red), PILOT_PA
6: LED2 (IR), PILOT_PA
7: LED3 (green), PILOT_PA
config MAX30101_SLOT4
int "Slot 4"
range 0 7
default 0
help
Set which LED and pulse amplitude are active in time slot 4.
0: None (disabled)
1: LED1 (red), LED1_PA
2: LED2 (IR), LED2_PA
3: LED3 (green), LED3_PA
4: None (disabled)
5: LED1 (red), PILOT_PA
6: LED2 (IR), PILOT_PA
7: LED3 (green), PILOT_PA
endif # MAX30101_MULTI_LED_MODE
endif # MAX30101